Abberior STED

The Abberior Facility Line FLIM STED microscope is an advanced imaging system that combines point scanning confocal microscopy with super-resolution and fluorescence lifetime imaging capabilities.

Abberior Facility Line STED System

The microscope uses Stimulated Emission Depletion (STED) technology to achieve higher spatial resolution beyond the diffraction limit of standard light microscopy(30nm in XY and 70nm in Z) . This makes it possible to visualize structures at the nanoscale. Additionally, the integration with Fluorescence Lifetime Imaging Microscopy (FLIM) allows for the measurement of the exponential decay rate of the fluorescence from a fluorescent sample, providing not just structural but also functional information about the molecules being observed. This combination of STED and FLIM opens up a range of possibilities for in-depth analysis in cellular biology, neurobiology, and materials science.

 

Photo of Abberior STED Microscope

STED vs Confocal

Comparison of Confocal(Top) and STED(Bottom) resolution.(Live cell mitochondria, scale bar 1 micron).

Specification

Imaging Laser:

405nm (cw, 20mW)

485nm (pulsed, 1 mW@ 40 MHz out of laser head)

561nm (pulsed, 0.2 mW@40 MHz out of laser head)

640nm (pulsed,1 mW@ 40 MHz out of laser head)

STED Laser:

775 nm > 2750mW @40Mhz

Easy3D for 775 nm STED

Objectives:

Objective NA WD Immersion Notes
UPLXAPO60XO 1.42 0.15mm Oil STED and Confocal Compatible
UPLSAPO60XS 1.3 0.3mm Silicon Oil

STED and Confocal Compatible

UPLAPOL60XW 1.2 0.2mm Water STED and Confocal Compatible
UPLXAPO40X 0.95 0.18mm Air Confocal Compatible
UPLXAPO20X 0.8 0.6mm Air Confocal Compatible
UPLXAPO10X 0.4 3.1mm Air Confocal Compatible

Detectors:

 

Transmitted Light Detector (PMT)

For fluorescence detection, there are two types of detectors:

a. Four Single-Photon-Counting APD Detectors thats compatible with Spectral Detection (QE around 65% at 500nm,low readout noise)

b. MATRIX Spectral APD array (FLIM, STED and Confocal Compatbile) (QE around 50% at 500nm,low readout noise)

Traditional PMT detectors on point scanning confocals have QE around 30% (Ref)

The MATRIX detector can also be used for Image Scanning Microscopy (same technique as the Zeiss Airyscan) and can achieve higher resolution(around 160nm) with regular fluorophores.
 

FLIM (Fluorescences Lifetime Imaging Microscopy)

The system can function as a single-detector FLIM (MATRIX Detector) that can be used to perform TCSPC based FLIM imaging (FLIM-STED, FLIM-Matrix(Imaging Scanning Microscopy, similar to Airyscan), FLIM-FRET) with temporal resolution of around 50 ps.

Common dye's life time plot (author: Dominik Frei)

Other Modules

FLEXPOSURE: Adaptive Illumination Package for Highest Resolution and Live cell
Imaging

ADAPTIVE OPTICS: Adaptive optics to correct optical distortion when imaging deeper into the sample.

Image Processing

The system is equipped with a Huygens Pro license for image deconvolution.

Sample Preparation

Special secondary antibody should be used for STED imaging (usually fluoresohpore that has some emission wavelength overlapping with the STED laser 775nm works). The NorMIC facility has an array of Abberior STAR secondary antibodies available for platform users to test. Please consult the platform staff to obtain an aliquot.

STED Sample Preparation Guide(Abberior)

STED Sample Preparation Guide(Leica)

Multi-color STED Imaging(Abberior)

5 Color STED antibody suggestion

Contact: xian.hu at medisin.uio.no

 